GAIA-FIELD Software Release Change Log
GAIA-FIELD Software Change Log
This document describes the software changes with each release of the GAIA-FIELD controller software.
Document Release
Version | Date | Description | Responsible |
1 | 2-Jan-2026 | Minor Release #4c5f856 | J Wingate |
1.1 | 17-Apr-2026 | Minor Release #9c027894 | J Wingate |
1.2 | 22-Apr-2026 | Minor Release #a5c147c6 | J Wingate |
1.3 | 19-May-2026 | Minor Release # ca2f3291 | J Wingate |
2026_05_19_gaia_software_update_ ca2f3291
New Features
- Alpha Correction Factor
- Alpha Correction Factor can be applied in Advanced Calibration Settings to reduce warm up non-linear drift. Contact SMG to determine your devices Alpha Factor (included as standard from June 2026 production units). See SMG Website Blog for further information.
- Hidden Advanced Calibration Settings
- Advanced calibration parameters are now hidden to reduce risk of accidental changes
- Live running gravity value
- During a reading the chart and table view displays the current gravity reading
- Improved Calibration Menu and Bias Calibration
- Neater calibration menu
- Easier to adjust Bias (the relative gravity value)
- Updated Home Page
- Cleaner home page user interface
Bug Fixes
- Fixed project load error after power up or return from standby
- Several minor bug fixes and UI improvements
Known Bugs
- None
2026_04_22_gaia_software_update_a5c147c6
Bug Fixes
- Sorting Station readings in Measure view does not function properly
- Fixed
Known Bugs
- After powerup or when returning from standby UI may falsely report project loading error initially.
2026_04_17_gaia_software_update_9c027894
Bug Fixes
- Temperature and Tide values in detailed station view
- Fixed the incorrect values in these fields
- Export column selection interface
- Fixed UI interface for column selection tool
- Current station plot
- Temperatures and correction values now show absolutes.
- Tablet / Phone layout
- Fixed issues for UI display on smaller screens
New Features
- Improved Project review and load screen, downloads page removed
- Live reading, stdev and sterr in table view while a measurement is being taken, stopping an active reading saves current values and data
- Start reading button becomes stop during reading
- Ability to update physics core firmware with each software update
- Improved UI for station name and station note edits
Known Bugs
- After powerup or when returning from standby UI may falsely report project loading error initially.
2026_01_02_gaia_software_update_4c5f856
Bug Fixes
- Drift calibration button not functioning
- Drift calibration button now automatically triggers an automated project to collect drift data for calibration. The drift project will ignore levelling errors that have previously stopped projects.
- Drift Calibration Result display errors
- Fix display of drift calibration results (previously if not calculating a temperature correction, the graphs would display incorrectly, causing the fit to appear incorrect)
- Calculating Drift Calibration Error
- Fixed error when running the drift calculation if the station was not called the default "drift".
- Project name duplication
- Fix implemented to check for duplicate project names (previously it would not work if the current month or day were < 10)
- Large project loading errors
- Fixed some issues when loading very large projects (with > 600 stations)
New Features
- NL (nonlinearity) corrections now implemented
Known Bugs
- Levelling Errors
- Levelling can trigger an error due to ‘timeout’ when performing larger levelling movements or in unstable conditions
§ Corrective action > Repeat station / levelling process